TECHNICAL,
SERVICE & CARE
74
®
TECHNICAL,
SERVICE & CARE
75
www.bayerdiabetes.com/us
Need Help? Call toll free: 1.800.348.8100 (available 24 hours a day, 7 days a week)
System Specifications
Test Sample: Capillary whole blood
Test Result: Referenced to plasma/serum glucose
Sample Volume: 0.6 μL
Measuring Range: 20–600 mg/dL
Countdown Time: 5 seconds
Memory: Stores most recent 1000 test results
Battery Type: Non-serviceable, 310mAh rechargeable
lithium polymer battery, 3.4v – 4.2v (5V input voltage)
Meter/Battery Life: 5 years
Charging Current: 500mA
Strip Storage Temperature Range: 41º – 86ºF
Control Storage Temperature Range: 48º – 86ºF
Meter Operating Temperature Range: 41º – 113ºF
Humidity: 10–93% RH
Dimensions: 3.8 in (H) x 1.2 in (W) x 0.8 in (T)
Weight: 48 grams
Sound: A beep sounds when the meter is turned
on or off, a test strip is inserted, when a test strip is
filled with blood, or when a test result appears on the
screen. Two beeps sound when a strip is removed
and the meter is turned off, or to alert the user. Twenty
beeps will sound when a programmed Reminder
sounds.
Electromagnetic Compatibility (EMC): The
C
ONTOUR
®
NEXT LINK meter complies with the
electromagnetic requirements specified in ISO 15197.
Electromagnetic emissions are low and unlikely to
interfere with other nearby electronic equipment, nor
are emissions from nearby electronic equipment likely
to interfere with the C
ONTOUR
®
NEXT LINK. Immunity to
electrostatic discharge meets the requirements of IEC
61000-4-2. It is advisable to avoid use of electronic
devices in very dry environments especially if synthetic
materials are present.
The CONTOUR NEXT LINK meter has been tested for
radio frequency interference at the frequency range
and test levels specified by ISO 15197. Immunity to
Radio Frequency meets the requirements of
IEC 61000-4-3. To avoid radio frequency interference
do not use the C
ONTOUR NEXT LINK meter near
cellular or cordless telephones, walkie talkies,
garage door openers, radio transmitters or other
electrical or electronic equipment that are sources of
electromagnetic radiation, as these may interfere with
the proper operation of the meter.
Caring for Your Meter
CAUTION: Only connect to a computer or charger
that is certified to UL 60950-1, 5V maximum.
A charging current of 500 mAh minimum is required.
If you use a cable, use the one provided in your
meter kit.
Store your meter in the carrying case provided
whenever possible.
Wash hands and dry them well before handling to
keep the meter and test strips free of oils and other
contaminants.
Handle the meter carefully to avoid damaging the
electronics or causing other malfunctions.
Caring for Your Meter
System Specifications
84359751_CntrNextLINK_UG_EN_FpBp_v4.indd 74-7584359751_CntrNextLINK_UG_EN_FpBp_v4.indd 74-75 3/29/12 12:19 PM3/29/12 12:19 PM